*Biological safety cabinet test
The test is carried out in accordance with EN 12469 and NSF 49 standards
*Chemical hoods test
The test is carried out in accordance with IS 1839 standard
*Carbon active-ductless fume hoods test
The test is carried out in accordance with NFX 15211 standard
*Laminar units test
The test is carried out in accordance with ISO 14644 or EU GMP ANNEX 1 standard
The test includes:
⦁ Test the flow velocity of the supply filter
⦁ Test the inflow velocity
⦁ Integrity test HEPA filters
⦁ Checking smoke flow directions using a smoke generator
⦁ Particle inspection – level of cleanliness in compliance with ISO 14644 or EU GMP ANNEX 1 standard
⦁ Noise testing
⦁ Test the light intensity measured in LUX
At the end of the tests the test results will be recorded in the test report including a serviceability tag in the hood.